Ultra-high-molecular-weight polyethylene (UHMWPE) has been used for over 40 years for acetabular cups in total hip joint prosthesis. Hylamer is a hot isostatically pressed material with high crystallinity. Early loosening due to extensive oxidation and phase transformation has been observed for this material. To analyze the reasons for its high tendency to oxidize, we have examined by FT-IR spectroscopy explanted acetabular cups γ-sterilized in different ways. The oxidation and wear rate observed were: Hylamer cups γ-sterilized in air with a long shelf life > Hylamer cups γ-sterilized in air with short shelf life Hylamer cups γ-sterilized in nitrogen. Our data indicate the important role of the shelf life on the oxidation of the cups: the samples γ-sterilized in air characterized by a shelf life greater than 3 years showed severe oxidation, a high wear rate, and marked debris production with delamination and the formation of a brittle zone in the more oxidated regions. In these regions we observed an increase in crystallinity, which was probably due to the decrease in the molecular weight of PE. The high oxidation was probably due to the modifications induced by the material treatment.

Vibrational spectroscopy study of the oxidation of Hylamer UHMWPE explanted acetabular cups sterilized differently / M. Reggiani; A. Tinti; M. Visentin; S. Stea; P. Erani; C. Fagnano. - In: JOURNAL OF MOLECULAR STRUCTURE. - ISSN 0022-2860. - STAMPA. - 834-836:(2007), pp. 129-135. [10.1016/j.molstruc.2006.11.007]

Vibrational spectroscopy study of the oxidation of Hylamer UHMWPE explanted acetabular cups sterilized differently.

REGGIANI, MATTEO;TINTI, ANNA;FAGNANO, CONCEZIO
2007

Abstract

Ultra-high-molecular-weight polyethylene (UHMWPE) has been used for over 40 years for acetabular cups in total hip joint prosthesis. Hylamer is a hot isostatically pressed material with high crystallinity. Early loosening due to extensive oxidation and phase transformation has been observed for this material. To analyze the reasons for its high tendency to oxidize, we have examined by FT-IR spectroscopy explanted acetabular cups γ-sterilized in different ways. The oxidation and wear rate observed were: Hylamer cups γ-sterilized in air with a long shelf life > Hylamer cups γ-sterilized in air with short shelf life Hylamer cups γ-sterilized in nitrogen. Our data indicate the important role of the shelf life on the oxidation of the cups: the samples γ-sterilized in air characterized by a shelf life greater than 3 years showed severe oxidation, a high wear rate, and marked debris production with delamination and the formation of a brittle zone in the more oxidated regions. In these regions we observed an increase in crystallinity, which was probably due to the decrease in the molecular weight of PE. The high oxidation was probably due to the modifications induced by the material treatment.
2007
Vibrational spectroscopy study of the oxidation of Hylamer UHMWPE explanted acetabular cups sterilized differently / M. Reggiani; A. Tinti; M. Visentin; S. Stea; P. Erani; C. Fagnano. - In: JOURNAL OF MOLECULAR STRUCTURE. - ISSN 0022-2860. - STAMPA. - 834-836:(2007), pp. 129-135. [10.1016/j.molstruc.2006.11.007]
M. Reggiani; A. Tinti; M. Visentin; S. Stea; P. Erani; C. Fagnano
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/48691
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 6
social impact