BACKGROUND: Metal-on-metal hip resurfacing arthroplasty (HR) has been gaining popularity especially for young and active patients. Although different series report good mid-term results, the long-term outcome and failure mechanisms are still concerning. In this consecutive revision case series, 9 retrieved specimens of a failed Birmingham Hip Resurfacing (BHR) were divided according to the time to fracture: 3 specimens failed at less than 6 months (Group 1), 3 failed between 6 months and 3 years (Group 2) and 3 failed later than 3 years (Group 3). The objective of the study was to examine by a specific quantitative histomorphometry and microtomography (micro-CT) method the characteristics of bone quality and its microarchitecture in retrieved metal-on-metal HR. METHODS: A series of 948 BHR were performed between 2001 and 2009. Among these implants 10 failures occurred and nine of these underwent revision surgery and were examined by histomorphometry and micro-CT. RESULTS: Histomorphometry showed a significant increase in trabecular separation (Tb.Sp) in Group 3 in comparison with Group 1 (113%, p < 0.05). In the top region, micro-CT showed that Groups 2 and 3 presented significant lower bone volume (Group 2: 61%, p < 0.005; Group 3: 1%, p < 0.05), trabecular number (Group 2: 53%, p < 0.005; Group 3: 40%, p < 0.05), and higher Tb.Sp (Group: 71%,p < 0.05) when compared to Group 1. Additionally, histomorphometry showed that the top regions in Group 1 had a significantly lower mean percentage of empty osteocyte lacunae than the top regions in both Group 2 and 3 (p < 0.05). CONCLUSIONS: This study showed that the morphometric parameters considered are crucial for a good understanding of mechanical properties of HR and may be of significant importance in the pathogenesis of HR failure particularly in the development of late fractures.

Histological, histomorphometric and microtomographic analyses of retrieval hip resurfacing arthroplasty failed at different times / Salamanna F.; Fini M.; Parrilli A.; Cadossi M.; Nicoli Aldini N.; Giavaresi G.; Luciani D.; Giannini S.. - In: BMC MUSCULOSKELETAL DISORDERS. - ISSN 1471-2474. - STAMPA. - 30;14(1):(2013), pp. 14-47. [10.1186/1471-2474-14-47]

Histological, histomorphometric and microtomographic analyses of retrieval hip resurfacing arthroplasty failed at different times.

CADOSSI, MATTEO;LUCIANI, DEIANIRA;GIANNINI, SANDRO
2013

Abstract

BACKGROUND: Metal-on-metal hip resurfacing arthroplasty (HR) has been gaining popularity especially for young and active patients. Although different series report good mid-term results, the long-term outcome and failure mechanisms are still concerning. In this consecutive revision case series, 9 retrieved specimens of a failed Birmingham Hip Resurfacing (BHR) were divided according to the time to fracture: 3 specimens failed at less than 6 months (Group 1), 3 failed between 6 months and 3 years (Group 2) and 3 failed later than 3 years (Group 3). The objective of the study was to examine by a specific quantitative histomorphometry and microtomography (micro-CT) method the characteristics of bone quality and its microarchitecture in retrieved metal-on-metal HR. METHODS: A series of 948 BHR were performed between 2001 and 2009. Among these implants 10 failures occurred and nine of these underwent revision surgery and were examined by histomorphometry and micro-CT. RESULTS: Histomorphometry showed a significant increase in trabecular separation (Tb.Sp) in Group 3 in comparison with Group 1 (113%, p < 0.05). In the top region, micro-CT showed that Groups 2 and 3 presented significant lower bone volume (Group 2: 61%, p < 0.005; Group 3: 1%, p < 0.05), trabecular number (Group 2: 53%, p < 0.005; Group 3: 40%, p < 0.05), and higher Tb.Sp (Group: 71%,p < 0.05) when compared to Group 1. Additionally, histomorphometry showed that the top regions in Group 1 had a significantly lower mean percentage of empty osteocyte lacunae than the top regions in both Group 2 and 3 (p < 0.05). CONCLUSIONS: This study showed that the morphometric parameters considered are crucial for a good understanding of mechanical properties of HR and may be of significant importance in the pathogenesis of HR failure particularly in the development of late fractures.
2013
Histological, histomorphometric and microtomographic analyses of retrieval hip resurfacing arthroplasty failed at different times / Salamanna F.; Fini M.; Parrilli A.; Cadossi M.; Nicoli Aldini N.; Giavaresi G.; Luciani D.; Giannini S.. - In: BMC MUSCULOSKELETAL DISORDERS. - ISSN 1471-2474. - STAMPA. - 30;14(1):(2013), pp. 14-47. [10.1186/1471-2474-14-47]
Salamanna F.; Fini M.; Parrilli A.; Cadossi M.; Nicoli Aldini N.; Giavaresi G.; Luciani D.; Giannini S.
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/133794
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 4
social impact