Abstract BACKGROUND: Relatively few data are available on long-term echocardiographic optimization of atrioventricular (AV) and interventricular (VV) delay programming in cardiac resynchronization therapy (CRT). We assessed variations in optimized AV and VV delays during long-term follow-up. METHODS: Thirty-seven consecutive heart failure patients received Doppler echocardiographic optimization of AV and VV delay within 48 hours from CRT device implantation, at 6 months and at 12 months (the last for the first enrolled 14 patients). RESULTS: After implantation, median optimized AV delay was 100 ms (range, 45 ms); VV optimization led to simultaneous biventricular activation in 4 patients, left ventricular preactivation in 17 patients and right ventricular preactivation in 16 patients. At 12 months median AV delay decreased to 85 ms (23 ms) (P < 0.05 vs. baseline). With respect to previous assessment, VV delay variations > or =40 ms were observed in 41% of the patients at 6 months and in 57% of the tested patients at 12 months. A nonconcordance (by Kappa test) of optimized VV delays was found between each new assessment and the previous one. VV delay optimization was associated with significant (P < 0.001) increases in aortic velocity time integral both at baseline and during follow-up. CONCLUSIONS: Echocardiographic optimization of AV and VV delay is associated with broad intraindividual variability during follow-up. A new assessment of optimized VV delays during long-term follow-up reveals a nonconcordance with previous values and provides increases in forward stroke volume.

Cardiac resynchronization therapy: variations in echo-guided optimized atrioventricular and interventricular delays during follow-up / Valzania C; Biffi M; Martignani C; Diemberger I; Bertini M; Ziacchi M; Bacchi L; Rocchi G; Rapezzi C; Branzi A; Boriani G.. - In: ECHOCARDIOGRAPHY. - ISSN 0742-2822. - STAMPA. - 24:9(2007), pp. 933-939. [10.1111/j.1540-8175.2007.00491.x]

Cardiac resynchronization therapy: variations in echo-guided optimized atrioventricular and interventricular delays during follow-up.

VALZANIA, CINZIA;BIFFI, MAURO;MARTIGNANI, CRISTIAN;DIEMBERGER, IGOR;BERTINI, MATTEO;ZIACCHI, MATTEO;BACCHI REGGIANI, MARIA LETIZIA;RAPEZZI, CLAUDIO;BRANZI, ANGELO;BORIANI, GIUSEPPE
2007

Abstract

Abstract BACKGROUND: Relatively few data are available on long-term echocardiographic optimization of atrioventricular (AV) and interventricular (VV) delay programming in cardiac resynchronization therapy (CRT). We assessed variations in optimized AV and VV delays during long-term follow-up. METHODS: Thirty-seven consecutive heart failure patients received Doppler echocardiographic optimization of AV and VV delay within 48 hours from CRT device implantation, at 6 months and at 12 months (the last for the first enrolled 14 patients). RESULTS: After implantation, median optimized AV delay was 100 ms (range, 45 ms); VV optimization led to simultaneous biventricular activation in 4 patients, left ventricular preactivation in 17 patients and right ventricular preactivation in 16 patients. At 12 months median AV delay decreased to 85 ms (23 ms) (P < 0.05 vs. baseline). With respect to previous assessment, VV delay variations > or =40 ms were observed in 41% of the patients at 6 months and in 57% of the tested patients at 12 months. A nonconcordance (by Kappa test) of optimized VV delays was found between each new assessment and the previous one. VV delay optimization was associated with significant (P < 0.001) increases in aortic velocity time integral both at baseline and during follow-up. CONCLUSIONS: Echocardiographic optimization of AV and VV delay is associated with broad intraindividual variability during follow-up. A new assessment of optimized VV delays during long-term follow-up reveals a nonconcordance with previous values and provides increases in forward stroke volume.
2007
Cardiac resynchronization therapy: variations in echo-guided optimized atrioventricular and interventricular delays during follow-up / Valzania C; Biffi M; Martignani C; Diemberger I; Bertini M; Ziacchi M; Bacchi L; Rocchi G; Rapezzi C; Branzi A; Boriani G.. - In: ECHOCARDIOGRAPHY. - ISSN 0742-2822. - STAMPA. - 24:9(2007), pp. 933-939. [10.1111/j.1540-8175.2007.00491.x]
Valzania C; Biffi M; Martignani C; Diemberger I; Bertini M; Ziacchi M; Bacchi L; Rocchi G; Rapezzi C; Branzi A; Boriani G.
File in questo prodotto:
Eventuali allegati, non sono esposti

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11585/116193
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? 8
  • Scopus 49
  • ???jsp.display-item.citation.isi??? 41
social impact